Hytrophobic magnetic beads protocol

Protocol

Protocol for diPhenyl Magnetic Beads

Products

1 μm BcMag™ diPhenyl Magnetic Beads
Cat. No.  FT105

Unit Size  250 mg
Order
1 μm BcMag™ diPhenyl Magnetic Beads
Cat. No.  FT106

Unit Size  500 mg
Order
5 μm BcMag™ diPhenyl Magnetic Beads
Cat. No.  FT107

Unit Size  250 mg
Order
5 μm BcMag™ diPhenyl Magnetic Beads
Cat. No.  FT108

Unit Size  500 mg
Order
Previous slide
Next slide

Cat. No.

FT105

FT106

FT107

FT108

Product Name

1 μm BcMag™ diPhenyl Magnetic Beads

1 μm BcMag™ diPhenyl Magnetic Beads

5 μm BcMag™ diPhenyl Magnetic Beads

5 μm BcMag™ diPhenyl Magnetic Beads

Unit Size

250 mg

500 mg

250 mg

500 mg

Specification

Composition

Magnetic beads grafted with diPhenyl groups

Number of Beads

~ 1.68 x 109 beads/mg (1μm beads)

~ 5 x 107 beads/mg (5μm beads)

Magnetization

~45 EMU/g

Type of Magnetization

Superparamagnetic

Effective Density

2.0 g/ml

Stability

Most organic solvents

Formulation

Lyophilized Powder

Binding Capacity

1 μm beads >20 μg protein/mg of Beads

5 μm beads >18 μg protein/mg of Beads

Storage

Store at 4°C upon receipt.

Introduction

Hydrophobic interaction chromatography (HIC) is an intricate technique utilized to segregate macromolecules based on their external hydrophobic region’s reversible interaction with hydrophobic ligands (such as phenyl, octyl, or butyl) of HIC medium. This process is influenced by the salt concentration of a buffer, where a high salt concentration enhances the interaction, and a low salt concentration reduces it. Consequently, based on salt concentration, the protein with lower hydrophobicity is eluted first, while the one with more hydrophobicity elutes last. This technique has gained immense popularity for protein and peptide purification in analytical and preparatory scale applications, as it employs a less denaturing environment.

The BcMag™ diPhenyl Magnetic Beads are highly uniform and superparamagnetic, with hydrophobic groups on their surface, making them ideal for chromatographic matrix applications. They can easily purify, desalinate, and concentrate peptides or proteins ranging from femtomolar to picomolar concentrations, eliminating the need for tedious pipetting and centrifugation procedures. These beads are particularly suitable for polypeptides with aromatic side chains, large hydrophobic proteins, membrane-spanning peptides, lipid peptides, and fusion proteins from inclusion bodies. Use BcMag™ diPhenyl Magnetic Beads for efficient and effective chromatography.

Structure of diPhenyl magnetic

PROTOCOL

Note:

To achieve maximum binding to the hydrophobic magnetic beads, TFA (trifluoroacetic acid) or other ion-pairing agents should be between 0.1%–1.0% at a pH of <4. The solvents should be completely removed if samples contain excess organic solvents such as methanol or acetonitrile (ACN). Samples can be dried in a vacuum evaporator and resuspended in sample buffer (below). To optimize binding, detergents in samples should be diluted with 0.1% TFA till SDS <0.1%, or Triton ® <1%, or Tween® <0.5%.

To avoid excessive beads drying between steps, the entire procedure should be carried out in a timely manner.

The amount of beads used in each application should be empirically titrated. The volumes can be scaled up or down accordingly. We recommend using 10 μl (0.5 mg) hydrophobic magnetic beads to bind ~ 10 μg protein and 5μl elution buffer for 0.5 mg beads.

Users are encouraged to determine the optimal working conditions based on the protocol and suggestions described in the Troubleshooting section to get the best results.

Materials Required

Buffers

Equilibration buffer: 0.5% TFA (trifluoroacetic acid) in 5% ACN (acetonitrile)

Sample Binding Buffer: 2% TFA in 5% ACN

Washing buffer: 0.5 % TFA in 5% ACN

Elution Buffer: 70% ACN

Magnetic Rack (for manual operation)

Based on sample volume, the user can choose one of the following Magnetic Racks:

– BcMag™ Magnetic Rack-2 for holding two individual 1.5 ml centrifuge tubes (Cat. No. MS-01);

– BcMag™ Magnetic Rack-6 for holding six individual 1.5 ml centrifuge tubes (Cat. No. MS-02);

– BcMag™ Magnetic Rack-24 for holding twenty-four individual 1.5-2.0 ml centrifuge tubes (Cat. No. MS-03);

– BcMag™ Magnetic Rack-50 for holding one 50 ml centrifuge tube, one 15 ml centrifuge tube, and four individual 1.5 ml centrifuge tubes (Cat. No. MS-04);

– BcMag™ Magnetic Rack-96 for holding a 96 ELISA plate or PCR plate (Cat. No. MS-05).

Procedure

A.

Magnetic Beads Preparation

1.

Weight and suspend 50 mg beads with 1ml of 50% methanol.

2.

Transfer 10μl (50 mg/ml) of completely suspended magnetic beads to a microcentrifuge tube.

3.

Place the tube onto a magnetic rack for 1-3 minutes until the supernatant is clear.

4.

Wash the beads with 1 ml of Equilibration buffer by a magnetic rack.

5.

Aspirate and discard the supernatant with a pipette while the tube remains in the rack.

B.

Sample Binding

1.

Mix sample (~10μg protein/ peptide) with 1/3 volume of Sample Binding Buffer and add to the tube containing the beads.

2.

Thoroughly mix beads and sample using a pipette and leave at room temperature for 2 minutes to allow proteins to bind to the beads.

3.

Place the tube onto the magnetic rack for 1-3 minutes (no longer than 3 minutes) until the supernatant is clear. Aspirate and discard the supernatant with a pipette while the tube remains in the rack.

4.

Remove the tube from the rack and resuspend the beads with 100μl washing buffer.

5.

Place the tube onto the magnetic rack for 1-3 minutes until the supernatant is clear. Aspirate and discard the supernatant with a pipette while the tube remains in the rack.

6.

Repeat steps 2 to 4 for four times.

C.

Elution

1.

Remove the tube from the rack, add 5μl elution buffer, resuspend the beads and incubate for 2 minutes at room temperature.

2.

Place the tube on the magnetic rack for 1-3 minutes and transfer the supernatant containing the eluted protein to a new tube. (User should optimize elution conditions for individual proteins by adjusting acetonitrile concentrations, such as 20%, 50%, 80%).

3.

For MALDI-MS analysis, mix 1μl of the eluate with 1μl of matrix solution and spot 0.5μl onto a MALADI-MS target plate.

Troubleshooting

Problem

Poor absorption of proteins/Peptides to beads.

Probable Cause

Hydrophobic interaction is not strong enough.

Suggestion

Increase the NaCl concentration (up to 0.2 M) used during adsorption.

Problem

Poor absorption of proteins/Peptides to beads.

Probable Cause

Biomolecules are not completely solubilized in the sample buffer.

Suggestion

Use denaturing conditions during adsorption. Add Guanidine HCl to the sample to achieve a final concentration between 1– 6 M.

Problem

Poor absorption of proteins/Peptides to beads.

Probable Cause

The sample’s chemical properties do not support hydrophobic interaction with reverse-phase beads.

Suggestion

Choose suitable reverse phase beads for your sample.

Problem

Poor elution

Probable Cause

Hydrophobic interaction is too strong.

Suggestion

Increase the acetonitrile concentration used during elution. Decrease the NaCl concentration used during adsorption.

Problem

Poor elution

Probable Cause

Proteins/peptides are not readily soluble in organic solutions.

Suggestion

Decrease the organic solvent concentration used during elution.

Problem

Poor elution

Probable Cause

Protein-bound too tightly to beads

Suggestion

Choose more suitable reverse phase beads for your sample.

Problem

Poor yield

Probable Cause

The quantity of protein or peptides of interest in the sample is too low.

Suggestion

  • If small sample quantities are used, decrease the amount of beads used and the volume of the elution buffer. An elution volume of 10μl acetonitrile per mg of beads is recommended.
  • Use a larger amount of starting sample.

Problem

Probable Causes

Suggestions

Poor absorption of proteins/Peptides to beads.

Hydrophobic interaction is not strong enough.

Increase the NaCl concentration (up to 0.2 M) used during adsorption.

Biomolecules are not completely solubilized in the sample buffer.

Use denaturing conditions during adsorption. Add Guanidine HCl to the sample to achieve a final concentration between 1– 6 M.

The sample’s chemical properties do not support hydrophobic interaction with reverse-phase beads.

Choose suitable reverse phase beads for your sample.

Poor elution

Hydrophobic interaction is too strong.

Increase the acetonitrile concentration used during elution. Decrease the NaCl concentration used during adsorption.

Proteins/peptides are not readily soluble in organic solutions.

Decrease the organic solvent concentration used during elution.

Protein-bound too tightly to beads

Choose more suitable reverse phase beads for your sample.

Poor yield

The quantity of protein or peptides of interest in the sample is too low.

  • If small sample quantities are used, decrease the amount of beads used and the volume of the elution buffer. An elution volume of 10μl acetonitrile per mg of beads is recommended.
  • Use a larger amount of starting sample.